GDHG-6802 transformer calibrator Product Description:
At present, the traditional transformer test methods are mostly used comparing the measured difference, that is, when the transformer during the test, will be seized transformer secondary connected load (load cases instead of commonly used), and standard transformer ratio is compared with, (booster) is provided to both the primary current (voltage) by the up-flow, and the second difference between the two flow (differential) input signal calibrator, the calibrator is measured relative to the standard mutual inductance transformer seized the error. This test method, a large current (voltage) power supplies, transformers and load box with standard variable ratio. When a current is greater than 1000A, a voltage higher than 110kV, the power supply capacity of up to tens of kVA, bulky and difficult wiring in the field test transformer inconvenient.
Instrument is based on a detailed analysis of the mathematical model developed on the transformer transformer tester. During transformer test, it does not require high-current power supplies and standard transformer and large load box, just to bring the instrument can be measured according to national standard differential current transformer ratio difference and angular, but also measuring the current 1% 120% of the difference between any percentage ratio and angle difference; direct measurement of current and voltage transformer turns ratio; worse than the qualitative difference and angle can be measured voltage transformer; field measurement of the actual current and voltage transformers secondary load, test particularly convenient. Using high-precision instrument, automatic linear regulator of DC and AC power supply; high-speed, high-reliability digital processing module; high accuracy, high stability of the pre-measurement circuit ensures high accuracy and stability of the measurement instrument. In addition, the instrument also has a user-friendly interface, and operating tips wiring, and save any measurement parameters and measurement results print function.

参数

GDHG-6802 transformer field calibrator for mutual verification, verifiable current transformers, voltage transformers. Compared with the transfer; when domestic traditional transformer test methods are mostly used comparing the measured difference, that is, during the test transformer, the transformer secondary subject connected load (instead of the usual load box), with the ratio of the standard transformer comparison, the up-flow (booster) is provided to both the primary current (voltage), while the secondary flow difference between the two (differential) input signal calibrator, the calibrator measured mutual inductance subject relative to the standard transformer errors. This test method, a large current (voltage) power supplies, transformers and load box with standard variable ratio. When a current is greater than 1000A, a voltage higher than 110kV, high power supply capacity. During transformer test, it does not require high-current power supplies and standard transformer and large load box, just to bring the instrument can be measured according to national standard differential current transformer ratio difference and angular, but also measuring the current 1% 120% of the difference between the percentage ratio of the difference and any angle. Can be directly measured current and voltage transformer turns ratio; worse than the qualitative difference and angle can be measured voltage transformer; secondary load can measure the actual field current and voltage transformers, test particularly convenient. Using high-precision instrument, automatic linear regulator of DC and AC power supply; high-speed, high-reliability digital processing module; high accuracy, high stability of the pre-measurement circuit ensures high accuracy and stability of the measurement instrument
 
Output voltage: 0 ~ 120V
Output current: 0 ~ 5Arms (15A peak)
Maximum knee voltage: 45000V
Ratio measuring range: 1 to 35,000
The maximum permissible error ratio measurement: ± 0.1%
Phase measuring the maximum permissible error: ± 3min
Secondary winding resistance measurement range: 0.05 ~ 200Ω
Secondary winding resistance measurement error: ± (0.1% × reading + 1mΩ)
Secondary load impedance measurement range: 0.05 ~ 200Ω
Secondary load impedance measurement error: ± (0.1% × reading + 1mΩ)
Input supply voltage: AC220V ± 10%, 50HZ
Ambient temperature: -10 ℃ ~ 50 ℃
Relative humidity: ≤90% non-condensing
Box Size: 320mm x364mm x 155mm (W x H x D)
Chassis weight: 11Kg
LCD: 8.4-inch color screen (visible in bright light)